Yesterday, reporter accompanied the province quality monitoring bureau measurement inspector to make secret inquiries the Fuzhou urban district 12 health clinics the sphygmomanometer quality situation, discovered only had 1 clinic the sphygmomanometer to surpass the examination cycle and appears the obvious data deviation. From this April, the entire province community clinic and the village level clinic's medical measurement appliance comprehensively accepts "the physical examination". North has the problem is located the main road Chinese rich clinic, two time the blood pressure which surveys for the province quality monitoring bureau measurement institute inspector is: Low pressure 58 mercury, high-pressured 80
mercury, but uses the blood pressure which the qualified sphygmomanometer determines, the low pressure value between 80 mercury, the high-pressured value between 120 mercury, the data has differed 20 ~ 40 mercury. The inspector discovered the Chinese rich clinic sphygmomanometer the examination certificate date of expiry is "in June, 2006", "the extended active duty" more than 1 year, was ordered to stop using. Province quality monitoring bureau measurement institute Director Chen tells reporter, the home use mercury sphygmomanometer must pay attention to the standard use, some instrument breakdowns are the user not when uses creates, like Sphygmomanometer each time uses does not crowd the armband in air cleanly, or the sphygmomanometer depositing temperature heat, has caused the mercury to divulge and so on